Scanning SQUID microscopy

Source 📝

677:
Nitrogen is used. During non-contact, non-destructive imaging of room temperature samples in air, the system achieves a raw, unprocessed spatial resolution equal to the distance separating the sensor from the current or the effective size of the sensor, whichever is larger. To best locate a wire short in a buried layer, however, a Fast Fourier Transform (FFT) back-evolution technique can be used to transform the magnetic field image into an equivalent map of the current in an integrated circuit or printed wiring board. The resulting current map can then be compared to a circuit diagram to determine the fault location. With this post-processing of a magnetic image and the low noise present in SQUID images, it is possible to enhance the spatial resolution by factors of 5 or more over the near-field limited magnetic image. The system's output is displayed as a false-color image of magnetic field strength or current magnitude (after processing) versus position on the sample. After processing to obtain current magnitude, this microscope has been successful at locating shorts in conductors to within ±16 μm at a sensor-current distance of 150 μm.
1324:
For purposes of this paper the top-down X-ray view shows the x-y plane of the module. The side view shows the x-z plane, and the end view shows the y-z plane. No anomalies were noted in the radiographic images. Excellent alignment of components on the mini-boards permitted an uncluttered top-down view of the mini-circuit boards. The internal construction of the module was seen to consist of eight, stacked mini-boards, each with a single microcircuit and capacitor. The mini-boards connected with the external module pins using the gold-plated exterior of the package. External inspection showed that laser-cut trenches created an external circuit on the device, which is used to enable, read, or write to any of the eight EEPROM devices in the encapsulated vertical stack. Regarding nomenclature, the laser-trenched gold panels on the exterior walls of the package were labeled with the pin numbers. The eight miniboards were labeled TSOP01 through TSOP08, beginning at the bottom of the package near the device pins.
1276:
superconducting sensors. Instruments equipped with such sensors can follow the path of a short circuit along its course through a part. The SQUID detector has been used in failure analysis for many years, and is now commercially available for use at the package level. The ability of SQUID to track the flow of current provides a virtual roadmap of the short, including the location in plan view of the shorting material in a package. We used the SQUID facilities at Neocera to investigate the failure in the package of interest, with pins carrying 1.47 milliamps at 2 volts. SQUID analysis of the part revealed a clear current path between the two pins of interest, including the location of the conductive material that bridged the two pins. The SQUID scan of the part is shown in Figure 1.
1176:
is only half the task; fault localization is still necessary. The critical step is to overlay the SSM current images or current path images with CAD files such as bonding diagrams or RTX images to pinpoint the fault location. To make alignment of overlaying possible, an optical two-point reference alignment is made. The package edge and package fiducial are the most convenient package markings to align to. Based on the data analysis, fault localization by SSM should isolate the short in the die, bond wires or package substrate. After all non-destructive approaches are exhausted, the final step is destructive deprocessing to verify SSM data. Depending on fault isolation, the deprocessing techniques include decapsulation, parallel lapping or cross-section.
1172:
analysis, and Real-Time X-ray (RTX) inspection were the non-destructive tools used to detect short faults. Unfortunately, these techniques do not work very well in advanced wire-bond packages. Because of the high density wire bonding in advanced wire-bond packages, it is extremely hard to localize the short with conventional RTX inspection. Without detailed information as to where the short might occur, attempting destructive decapsulation to expose both die surface and bond wires is full of risk. Wet chemical etching to remove mold compound in a large area often results in over-etching. Furthermore, even if the package is successfully decapped, visual inspection of the multi-tiered bond wires is a blind search.
1328:
wide trace on the mini-boards. All of the Vss pins were shorted to the Vcc pins with a resistance determined by the I-V slope at approximately 1.74 ohms, the low resistance indicating something other than an ESD defect. Similarly electrical overstress was considered an unlikely cause of failure as the part had not been under power since the time it was qualified at the factory. The three-dimensional geometry of the EEPROM module suggested the use of magnetic current imaging (MCI) on three, or more flat sides in order to construct the current path of the short within the module. As noted, the coordinate axes selected for this analysis are shown in Figure 1.
1228:
of the package. Based on electrical and layout analysis of the package, it can be inferred that current is either flowing through the wirebond pads or that the wirebonds are somehow touching a conductive structure at the specified location. After obtaining similar SSM results on the two units under test, further destructive analysis focused around the small potential short region, and it showed that the failing pin wirebond is touching the bottom of one of the stacked dice at the specific XY position highlighted by SSM analysis. The cross section view of one of those units is shown in figure 3.
1156: 1212: 1204: 1196: 745:. When the SQUID is biased with a constant current that exceeds the critical current of the junction, then changes in the magnetic flux, Φ, threading the SQUID loop produce changes in the voltage drop across the SQUID (see Figure 1). Figure 2(a) shows the I-V characteristic of a SQUID where ∆V is the modulation depth of the SQUID due to external magnetic fields. The voltage across a SQUID is a nonlinear periodic function of the applied magnetic field, with a periodicity of one flux quantum, Φ 1337:
wide), although software and data acquisition improvements allow locating currents within 3 micrometres. To operate, the SQUID sensor must be kept cool (about 77 K) and in vacuum, while the sample, at room temperature, is raster-scanned under the sensor at some working distance z, separated from the SQUID enclosure by a thin, transparent diamond window. This allows one to reduce the scanning distance to tens of micrometres from the sensor itself, improving the resolution of the tool.
20: 1259: 1251: 493:. To use the DC SQUID to measure standard magnetic fields, one must either count the number of oscillations in the voltage as the field is changed, which is very difficult in practice, or use a separate DC bias magnetic field parallel to the device to maintain a constant voltage and consequently constant magnetic flux through the loop. The strength of the field being measured will then be equal to the strength of the bias magnetic field passing through the SQUID. 1285:
point that radiography can now be used to identify features heretofore impossible to detect. The equipment at Xradia was used to inspect the failure of interest in this analysis. An example of their findings is shown in Figure 2. The feature shown (which is also the material responsible for the failure) is a copper filament approximately three micrometres wide in cross-section, which was impossible to resolve in our in-house radiography equipment.
749:=2.07×10 Tm (see Figure 2(b)). In order to convert this nonlinear response to a linear response, a negative feedback circuit is used to apply a feedback flux to the SQUID so as to keep the total flux through the SQUID constant. In such a flux locked loop, the magnitude of this feedback flux is proportional to the external magnetic field applied to the SQUID. Further description of the physics of SQUIDs and SQUID microscopy can be found elsewhere. 1220:
defect locations; however, defects like metal migration produced at wirebond pads, or bond wires somehow touching any other conductive structures, may be very difficult to catch with non-destructive techniques that are not electrical in nature. Here, the availability of analytical tools that can map out the flow of electric current inside the package provide valuable information to guide the failure analyst to potential defect locations.
741:, cooled below 80K and in vacuum while the device under test is at room temperature and in air. A SQUID consists of two Josephson tunnel junctions that are connected together in a superconducting loop (see Figure 1). A Josephson junction is formed by two superconducting regions that are separated by a thin insulating barrier. Current exists in the junction without any voltage drop, up to a maximum value, called the critical current, I 657: 2137: 1267:
with conventional radiographic analysis were unsuccessful. Arguably the most difficult part of the procedure is identifying the physical location of the short with a high enough degree of confidence to permit destructive techniques to be used to reveal the shorting material. Fortunately, two analytical techniques are now available that can significantly increase the effectiveness of the fault localization process.
891: 1289:
both time and money to perform. In effect, to get the most out of it, the analyst really needs to know already where the failure is located. This makes low-power radiography a useful supplement to SQUID, but not a generally effective replacement for it. It would likely best be used immediately after SQUID to characterize morphology and depth of the shorting material once SQUID had pinpointed its location.
104: 686: 702: 1164: 1315: 1307: 1298: 426: 1224:
inconsistently failing and recovering under reliability tests. Time domain reflectometry and X-ray analysis were performed on these units with no success in isolating the defects. Also there was no clear indication of defects that could potentially produce the observed electrical short failure mode. Two of those units were analyzed with SSM.
1341:
another, getting closer to the sensor in the process, this will be revealed as stronger magnetic field intensity for the section closer to the sensor and also as higher intensity in the current density map. This way, vertical information can be extracted from the current density images. Further details about MCI can be found elsewhere.
644:. Depending on the particular application, different levels of precision may be required in the height of the apparatus. Operating at lower-tip sample distances increases the sensitivity and resolution of the device, but requires more advanced mechanisms in controlling the height of the probe. In addition such devices require extensive 1340:
The typical MCI sensor configuration is sensitive to magnetic fields in the perpendicular z direction (i.e., sensitive to the in-plane xy current distribution in the DUT). This does not mean that we are missing vertical information; in the simplest situation, if a current path jumps from one plane to
1327:
Pin-to-pin electrical testing confirmed that Vcc Pins 12, 13, 14, and 15 were electrically common, presumably through the common exterior gold panel on the package wall. Likewise, Vss Pins 24, 25, 26, and 27 were common. Comparison to the xray images showed that these four pins funneled into a single
1058:
of a single YBCO crystal (figure). In a Josephson junction ring the superconducting electrons form a coherent wave function, just as in a superconductor. As the wavefunction must have only one value at each point, the overall phase factor obtained after traversing the entire Josephson circuit must be
1544:
L. A. Knauss, B. M. Frazier, H. M. Christen, S. D. Silliman and K. S. Harshavardhan, Neocera LLC, 10000 Virginia Manor Rd. Beltsville, MD 20705, E. F. Fleet and F. C. Wellstood, Center for Superconductivity Research, University of Maryland at College Park College Park, MD 20742, M. Mahanpour and A.
1227:
Electrically connecting the failing pin to a ground pin produced the electric current path shown in figure 2. This electrical path strongly suggests that the current is somehow flowing through all the ground nets though a conductive path located very close to the wirebond pads from the top down view
881:
As a result, the current can be directly calculated from the magnetic field knowing only the separation between the current and the magnetic field sensor. The details of this mathematical calculation can be found elsewhere, but what is important to know here is that this is a direct calculation that
676:
of current at a distance of 100 μm from the SQUID sensor with 1 second averaging. The microscope uses a patented design to allow the sample under investigation to be at room temperature and in air while the SQUID sensor is under vacuum and cooled to less than 80 K using a cryo cooler. No Liquid
1323:
Examination of the module shown in Figure 1 in the Failure Analysis Laboratory found no external evidence of the failure. Coordinate axes of the device were chosen as shown in Figure 1. Radiography was performed on the module in three orthogonal views: side, end, and top-down; as shown in Figure 2.
1288:
The principal drawback of this technique is that the depth of field is extremely short, requiring many ‘cuts’ on a given specimen to detect very small particles or filaments. At the high magnification required to resolve micrometre-sized features, the technique can become prohibitively expensive in
1266:
The failure in this example was characterized as an eight-ohm short between two adjacent pins. The bond wires to the pins of interest were cut with no effect on the short as measured at the external pins, indicating that the short was present in the package. Initial attempts to identify the failure
1175:
The Scanning SQUID Microscopy (SSM) data are current density images and current peak images. The current density images give the magnitude of the current, while the current peak images reveal the current path with a ± 3 μm resolution. Obtaining the SSM data from scanning advanced wire-bond packages
1171:
Advanced wire-bond packages, unlike traditional Ball Grid Array (BGA) packages, have multiple pad rows on the die and multiple tiers on the substrate. This package technology has brought new challenges to failure analysis. To date, Scanning Acoustic Microscopy (SAM), Time Domain Reflectometry (TDR)
619:
should be chosen. The SQUID itself can be used as the pickup coil for measuring the magnetic field, in which case the resolution of the device is proportional to the size of the SQUID. However, currents in or near the SQUID generate magnetic fields which are then registered in the coil and can be a
724:
the tip across the area where measurements are desired. As the change in voltage corresponding to the measured magnetic field is quite rapid, the strength of the bias magnetic field is typically controlled by feedback electronics. This field strength is then recorded by a computer system that also
1284:
The second fault location technique will be taken somewhat out of turn, as it was used to characterize this failure after the SQUID analysis, as an evaluation sample for an equipment vendor. The ability to focus and resolve low-power X-rays and detect their presence or absence has improved to the
1219:
Electric shorts in multi-stacked die packages can be very difficult to isolate non-destructively; especially when a large number of bond wires are somehow shorted. For instance, when an electric short is produced by two bond wires touching each other, X-ray analysis may help to identify potential
628:
of the device, and limitations in the control of the bias magnetic field as well as electronics issues prevent a perfectly constant voltage from being maintained at all times. However, in practice, the sensitivity in most scanning SQUID microscopes is sufficient for almost any SQUID size for many
1336:
SQUIDs are the most sensitive magnetic sensors known. This allows one to scan currents of 500 nA at a working distance of about 400 micrometres. As for all near field situations, the resolution is limited by the scanning distance or, ultimately, by the sensor size (typical SQUIDs are about 30 μm
1223:
Figure 1a shows the schematic of our first case study consisting of a triple-stacked die package. The X-ray image of figure 1b is intended to illustrate the challenge of finding the potential short locations represented for failure analysts. In particular, this is one of a set of units that were
1275:
One characteristic that all shorts have in common is the movement of electrons from a high potential to a lower one. This physical movement of the electrical charge creates a small magnetic field around the electron. With enough electrons moving, the aggregate magnetic field can be detected by
728:
As the name implies, SQUIDs are made from superconducting material. As a result, they need to be cooled to cryogenic temperatures of less than 90 K (liquid nitrogen temperatures) for high temperature SQUIDs and less than 9 K (liquid helium temperatures) for low temperature SQUIDs. For magnetic
620:
source of noise. To reduce this effect it is also possible to make the size of the SQUID itself very small, but attach the device to a larger external superconducting loop located far from the SQUID. The flux through the loop will then be detected and measured, inducing a voltage in the SQUID.
1982:
Wills, K.S., Diaz de Leon, O., Ramanujachar, K., and Todd, C., “Super-conducting Quantum Interference Device Technique: 3-D Localization of a Short within a Flip Chip Assembly,” Proceedings of the 27th International Symposium for Testing and Failure Analysis, San Jose, CA, November, 2001, pp.
504:, is also emitted in the bias coil. This AC field produces an AC voltage with amplitude proportional to the DC component in the SQUID. The advantage of this technique is that the frequency of the voltage signal can be chosen to be far away from that of any potential noise sources. By using a 227: 1992:"Construction of a 3-D Current Path Using Magnetic Current Imaging", ISTFA 2007, Frederick Felt, NASA Goddard Space Flight Center, Greenbelt, MD, USA, Lee Knauss, Neocera, Beltsville, MD, USA, Anders Gilbertson, Neocera, Beltsville, MD, USA, Antonio Orozco, Neocera, Beltsville, MD, USA 1973:"A Procedure for Identifying the Failure Mechanism Responsible for A Pin-To-Pin Short Within Plastic Mold Compound Integrated Circuit Packages", ISTFA 2008, Carl Nail, Jesus Rocha, and Lawrence Wong National Semiconductor Corporation, Santa Clara, California, United States 1078:)π will be observed. Compared to the case of standard s-wave junctions, where no phase shift is observed, no anomalous effects were expected in the B, C, and D cases, as the single valued property is conserved, but for device A, the system must do something to for the φ=2 757:
Magnetic current imaging uses the magnetic fields produced by currents in electronic devices to obtain images of those currents. This is accomplished through the fundamental physics relationship between magnetic fields and current, the Biot-Savart Law:
869: 882:
is not influenced by other materials or effects, and that through the use of Fast Fourier Transforms these calculations can be performed very quickly. A magnetic field image can be converted to a current density image in about 1 or 2 seconds.
488:
with respect to the amount of magnetic flux passing through the device. As a result, alone a SQUID can only be used to measure the change in magnetic field from some known value, unless the magnetic field or device size is very small such that
1318:
Figure 3: A magnetic current image overlay on an X-ray image of the EEPROM module. Thresholding was used to show only the strongest current at the capacitor of the TSOP08 mini-board. Arrows indicate Vcc and Vss pins. This image is in the x-y
1082:π condition to be maintained. In the same property behind the scanning SQUID microscope, the phase of the wavefunction is also altered by the amount of magnetic flux passing through the junction, following the relationship Δφ=π(Φ 421:{\displaystyle {\begin{aligned}V&={\frac {R}{2}}{\sqrt {I^{2}-I_{0}^{2}}},\\&={\frac {R}{2}}\left(I^{2}-\left(2I_{c}\cos \left(\pi {\frac {\Phi }{\Phi _{0}}}\right)\right)^{2}\right)^{\frac {1}{2}},\end{aligned}}} 623:
The resolution and sensitivity of the device are both proportional to the size of the SQUID. A smaller device will have greater resolution but less sensitivity. The change in voltage induced is proportional to the
1029: 1301:
Figure 1: An external view of the EEPROM module shows the coordinate axis used while performing orthogonal magnetic current imaging. These axes are used to define the scanning planes in the body of the
532:
can map out buried current-carrying wires by measuring the magnetic fields produced by the currents, or can be used to image fields produced by magnetic materials. By mapping out the current in an
1789:
Tsuei, C.C.; J. R. Kirtley; C. C. Chi; Lock See Yu-Jahnes; A. Gupta; T. Shaw; J. Z. Sun; M. B. Ketchen (1994). "Pairing Symmetry and Flux Quantization in a Tricrystal Superconducting Ring of YBa
232: 1526:
J. P. Wikswo, Jr. “The Magnetic Inverse Problem for NDE”, in H. Weinstock (ed.), SQUID Sensors: Fundamentals, Fabrication, and Applications, Kluwer Academic Publishers, pp. 629-695, (1996)
1062:
In YBCO, upon crossing the planes in momentum (and real) space, the wavefunction will undergo a phase shift of π. Hence if one forms a Josephson ring device where this plane is crossed (2
1112:
Scanning SQUID Microscope can detect all types of shorts and conductive paths including Resistive Opens (RO) defects such as cracked or voided bumps, Delaminated Vias, Cracked traces/
1109:
was the area of the ring. The device observed zero field at B, C, and D. The results provided one of the earliest and most direct experimental confirmations of d-wave pairing in YBCO.
901:
The scanning SQUID microscope was originally developed for an experiment to test the pairing symmetry of the high-temperature cuprate superconductor YBCO. Standard superconductors are
1031:, meaning that when crossing over any of the directions in momentum space one will observe a sign change in the order parameter. The form of this function is equal to that of the 75:, the scanning SQUID microscope allows magnetic fields to be measured with unparalleled resolution and sensitivity. The first scanning SQUID microscope was built in 1992 by Black 1452:
Black, R.C.; A. Mathai; and F. C. Wellstood; E. Dantsker; A. H. Miklich; D. T. Nemeth; J. J. Kingston; J. Clarke (1993). "Magnetic microscopy using a liquid nitrogen cooled YBa
566:
squid and maintaining thermal separation with the sample. In either case, due to the extreme sensitivity of the SQUID probe to stray magnetic fields, in general some form of
1097:
used a scanning SQUID microscope to measure the local magnetic field at each of the devices in the figure, and observed a field in ring A approximately equal in magnitude Φ
764: 1964:"Scanning SQUID Microscopy for New Package Technologies", ISTFA 2004, Mario Pacheco and Zhiyong Wang Intel Corporation, 5000 W. Chandler Blvd., Chandler, AZ, U.S.A., 85226 71:
near the surface of the sample to be measured. As the SQUID is the most sensitive detector of magnetic fields available and can be constructed at submicrometre widths via
729:
current imaging systems, a small (about 30 μm wide) high temperature SQUID is used. This system has been designed to keep a high temperature SQUID, made from YBa
633:
techniques it is possible to fabricate devices with total area of 1–10 μm, although devices in the tens to hundreds of square micrometres are more common.
199: 179: 152: 1059:
an integer multiple of 2π, as otherwise, one would obtain a different value of the probability density depending on the number of times one traversed the ring.
923: 125: 500:
technique is used. In addition to the DC bias magnetic field, an AC magnetic field of constant amplitude, with field strength generating Φ << Φ
1039:
function, giving it the name d-wave superconductivity. As the superconducting electrons are described by a single coherent wavefunction, proportional to exp(-
1121: 1923: 496:
Although it is possible to read the DC voltage between the two terminals of the SQUID directly, because noise tends to be a problem in DC measurements, an
2328: 2064: 544: 1879:
Sood, Bhanu; Pecht, Michael (2011-08-11). "Conductive filament formation in printed circuit boards: effects of reflow conditions and flame retardants".
1564:
Fleet, E.F.; Chatraphorn, S.; Wellstood, F.C.; Green, S.M.; Knauss, L.A. (1999). "HTS scanning SQUID microscope cooled by a closed-cycle refrigerator".
552: 1199:
Figure 1(a) Schematic showing typical bond wires in a triple-stacked die package, Figure 1(b) X-ray lateral view of actual triple-stacked die package.
640:
and operated either in direct contact with or just above the sample surface. The position of the SQUID is usually controlled by some form of electric
2001:
L. A. Knauss et al., "Current Imaging using Magnetic Field Sensors". Microelectronics Failure Analysis Desk Reference 5th Ed., pages 303-311 (2004).
1086:). As was predicted by Sigrist and Rice, the phase condition can then be maintained in the junction by a spontaneous flux in the junction of value Φ 1667:
Chatraphorn, S.; Fleet, E. F.; Wellstood, F. C.; Knauss, L. A.; Eiles, T. M. (17 April 2000). "Scanning SQUID microscopy of integrated circuits".
1262:
Figure 2: High-resolution radiographic image of filament, measured at 2.9 micrometres wide. Image shows filament running under both shorted leads.
539:
As the SQUID material must be superconducting, measurements must be performed at low temperatures. Typically, experiments are carried out below
2119: 725:
keeps track of the position of the probe. An optical camera can also be used to track the position of the SQUID with respect to the sample.
2124: 1113: 52: 936: 2107: 2057: 713:
respectively. Figure 2 b) Periodic voltage response due to flux through a SQUID. The periodicity is equal to one flux quantum, Φ
2184: 2154: 536:
or a package, short circuits can be localized and chip designs can be verified to see that current is flowing where expected.
1844:
Sigrist, Manfred; T. M. Rice (1992). "Paramagnetic Effect in High T c Superconductors -A Hint for d-Wave Superconductivity".
1207:
Figure 2: Overlay of current density, optical, and CAD images in triple-stacked die package with electric short failure mode.
933:
will be the same. However, in the high-temperature cuprate superconductors, the order parameter instead follows the equation
2236: 2231: 2216: 2179: 2102: 1395:
Shibata, Yusuke; Nomura, Shintaro; Kashiwaya, Hiromi; Kashiwaya, Satoshi; Ishiguro, Ryosuke; Takayanagi, Hideaki (2015).
2364: 2092: 2050: 664:
A high temperature Scanning SQUID Microscope using a YBCO SQUID is capable of measuring magnetic fields as small as 20
1508: 697:
is the critical current of the SQUID, Φ is the flux threading the SQUID and V is the voltage response to that flux.
181:. The thin barriers on each path are Josephson junctions, which together separate the two superconducting regions. 84: 508:
the device can read only the frequency corresponding to the magnetic field, ignoring many other sources of noise.
214: 2246: 2211: 2023: 1948: 1535:
E.F. Fleet et al., “HTS Scanning SQUID Microscopy of Active Circuits”, Appl. Superconductivity Conference (1998)
2354: 2226: 2206: 2201: 2164: 1310:
Figure 2: Radiography, showing three orthogonal views of the part, reveals internal construction of the module.
629:
applications, and therefore the tendency is to make the SQUID as small as possible to enhance resolution. Via
2169: 2114: 1503:. Vol. I: Fundamentals and Technology of SQUIDs and SQUID Systems. Weinheim: Wiley-VCH. pp. 46–48. 1047:
of the wavefunction, this property can be also interpreted as a phase shift of π under a 90 degree rotation.
80: 1925:
Failure Analysis of Short Faults on Advanced Wire-bond and Flip-chip Packages with Scanning SQUID Microscopy
559:
cooling to instead be used. It is even possible to measure room-temperature samples by only cooling a high
2136: 574:, possibly in combination with a superconducting "can" (all superconductors repel magnetic fields via the 2266: 2261: 1624:
Wellstood, F.C.; Gim, Y.; Amar, A.; Black, R.C.; Mathai, A. (1997). "Magnetic microscopy using SQUIDs".
2369: 2282: 2196: 1215:
Figure 3: Cross sectional image showing a bond wire touching the die causing signal to ground leakage.
864:{\displaystyle d{\vec {B}}={\frac {\mu _{0}}{4\pi }}{\frac {Id{\vec {l}}\times {\vec {r}}}{r^{3}}}\,.} 209:
SQUID. A DC SQUID consists of superconducting electrodes in a ring pattern connected by two weak-link
2349: 2292: 2251: 2174: 2159: 2073: 1710:
Wells, Frederick S.; Pan, Alexey V.; Wang, X. Renshaw; Fedoseev, Sergey A.; Hilgenkamp, Hans (2015).
669: 1167:
Optical image of the decapped wire bonds that are lifted from the die and touching another wire bond
1132:(MCM) and stacked die. SQUID scanning can also isolate defective components in assembled devices or 1116:
and Cracked Plated Through Holes (PTH). It can map power distributions in packages as well as in 3D
615:
superconductors. In either case, a superconductor with critical temperature higher than that of the
2221: 720:
Operation of a scanning SQUID microscope consists of simply cooling down the probe and sample, and
44: 2032: 905:
with respect to their superconducting properties, that is, for any direction of electron momentum
2287: 2087: 672:). The SQUID sensor is sensitive enough that it can detect a wire even if it is carrying only 10 1921: 1163: 1155: 2241: 1397:"Imaging of current density distributions with a Nb weak-link scanning nano-SQUID microscope" 1133: 616: 548: 474: 436: 877:
is the permeability of free space, and r is the distance between the current and the sensor.
705:
Figure 2 a) Plot of current vs. voltage for a SQUID. Upper and lower curves correspond to nΦ
1853: 1810: 1745: 1676: 1633: 1573: 1473: 1408: 582: 184: 157: 130: 8: 2359: 1499:
Boris Chesca; Reinhold Kleiner; Dieter Koelle (2004). J. Clarke; A. I. Braginski (eds.).
497: 210: 36: 1857: 1814: 1749: 1680: 1637: 1577: 1477: 1412: 1940: 1904: 1766: 1735: 1711: 1597: 1429: 1396: 1117: 1036: 908: 630: 589:. A wide variety of superconducting materials can be used, but the two most common are 567: 533: 110: 1908: 1896: 1826: 1771: 1692: 1649: 1589: 1504: 1434: 1129: 1125: 505: 1601: 1554:"Current Imaging using Magnetic Field Sensors" L.A. Knauss, S.I. Woods and A. Orozco 2297: 1944: 1936: 1932: 1888: 1861: 1818: 1761: 1753: 1684: 1641: 1581: 1481: 1424: 1416: 1370: 1350: 444: 64: 2027: 1139: 926: 608: > 77 K and relative ease of deposition compared to other high 594: 575: 556: 1270: 484:. However, as noted by the figure, the voltage across the electrodes oscillates 2313: 1822: 1211: 1203: 1195: 1159:
Current Image overlaid the optical image of the part and the layout of the part
894: 641: 521: 206: 56: 1892: 2343: 1900: 1788: 1712:"Analysis of low-field isotropic vortex glass containing vortex groups in YBa 1696: 1653: 1593: 1375: 1044: 873:
B is the magnetic induction, Idℓ is an element of the current, the constant μ
721: 540: 466: 68: 32: 2042: 1451: 19: 1922:
Steve K. Hsiung; Kevan V. Tan; Andrew J. Komrowski; Daniel J. D. Sullivan.
1830: 1775: 1438: 1258: 1250: 665: 455: 1632:(2). Institute of Electrical and Electronics Engineers (IEEE): 3134–3138. 1572:(2). Institute of Electrical and Electronics Engineers (IEEE): 3704–3707. 1074:, or even number of crossings, as in B, C, and D, a phase difference of (2 1054:
by manufacturing a series of YBCO ring Josephson junctions which crossed
1865: 1498: 1055: 586: 72: 28: 2323: 2318: 2020: 1545:
Ghaemmaghami, Advanced Micro Devices, One AMD Place Sunnyvale, CA 94088
1360: 1355: 930: 637: 625: 529: 481: 60: 27:
refrigerator. Green holder for the SQUID probe is attached to a quartz
1757: 1645: 1585: 1420: 201:
represents the magnetic flux entering the inside of the DC SQUID loop.
2189: 1688: 1485: 902: 645: 520:
is a sensitive near-field imaging system for the measurement of weak
2014: 1740: 656: 571: 465:
is the critical current of the Josephson junctions, Φ is the total
39:
of a SQUID probe and a test image of Nb/Au strips recorded with it.
24: 1234: 590: 218: 890: 2097: 1666: 1394: 673: 1563: 1140:
Short Localization in Advanced Wirebond Semiconductor Package
2033:
Center for Superconductivity Research, University of Maryland
1365: 1271:
Superconducting Quantum Interference Device (SQUID) Detection
1254:
Figure 1 SQUID image of package indicating location of short.
651: 525: 92: 1314: 1306: 1297: 1024:{\displaystyle \Delta (k)=\Delta _{0}(cos(k_{x}a)-(k_{y}a))} 701: 685: 598: 485: 103: 88: 2037: 205:
The scanning SQUID microscope is based upon the thin-film
648:
dampening if precise height control is to be maintained.
524:
by moving a Superconducting Quantum Interference Device (
127:
enters and splits into the two paths, each with currents
217:
of the Josephson junctions, the idealized difference in
752: 593:, due to its relatively good resistance to damage from 551:. However, advances in high-temperature superconductor 16:
Method of imaging magnetic fields at microscopic scales
2021:
Design and applications of a scanning SQUID microscope
1881:
Journal of Materials Science: Materials in Electronics
1070:+1)π will be observed between the two junctions. For 2 1623: 939: 911: 767: 230: 187: 160: 133: 113: 1709: 23:
Left: Schematic of a scanning SQUID microscope in a
2017:, one of the pioneers in scanning SQUID microscopy. 1724:
thin films visualized by scanning SQUID microscopy"
1179: 480:Hence, a DC SQUID can be used as a flux-to-voltage 67:. A tiny SQUID is mounted onto a tip which is then 1023: 917: 863: 420: 193: 173: 146: 119: 79:Since then the technique has been used to confirm 1843: 689:Figure 1: Electrical schematic of a SQUID where I 2341: 1235:Short between pins in molding compound package 1231:A similar defect was found in the second unit. 897:in YBCO imaged by the scanning SQUID microscopy 1626:IEEE Transactions on Applied Superconductivity 1566:IEEE Transactions on Applied Superconductivity 1464:superconducting quantum interference device". 1066:+1), number of times, a phase difference of (2 2072: 2058: 581:The actual SQUID probe is generally made via 1614:J. Kirtley, IEEE Spectrum p. 40, Dec. (1996) 925:in the superconductor, the magnitude of the 1331: 53:superconducting quantum interference device 2065: 2051: 652:High temperature scanning SQUID microscope 1878: 1765: 1739: 1428: 857: 570:is used. Most common is a shield made of 1492: 1313: 1305: 1296: 1292: 1279: 1257: 1249: 1210: 1202: 1194: 1162: 1154: 889: 700: 684: 655: 102: 18: 1522: 1520: 668:(about 2 million times weaker than the 98: 2342: 2140:Typical atomic force microscopy set-up 2046: 1050:This property was exploited by Tsuei 929:and consequently the superconducting 1517: 753:Magnetic field detection using SQUID 555:have allowed relatively inexpensive 636:The SQUID itself is mounted onto a 221:between the electrodes is given by 107:Diagram of a DC SQUID. The current 13: 2135: 956: 940: 511: 368: 364: 188: 14: 2381: 2008: 1675:(16). AIP Publishing: 2304–2306. 585:with the SQUID area outlined via 55:(SQUID) is used to image surface 1180:Short in multi-stacked packages 85:high-temperature superconductors 81:unconventional superconductivity 2247:Scanning quantum dot microscopy 1995: 1986: 1976: 1967: 1958: 1915: 1872: 1837: 1782: 1703: 1660: 1617: 885: 2202:Photothermal microspectroscopy 1937:10.31399/asm.cp.istfa2004p0073 1608: 1557: 1548: 1538: 1529: 1445: 1388: 1018: 1015: 999: 993: 977: 965: 949: 943: 838: 823: 777: 543:temperature (4.2 K) in a 1: 1381: 1043:φ), where φ is known as the 680: 7: 2185:Near-field scanning optical 2155:Ballistic electron emission 1344: 10: 2386: 2283:Scanning probe lithography 1823:10.1103/PhysRevLett.73.593 2365:Scanning probe microscopy 2306: 2293:Feature-oriented scanning 2275: 2257:Scanning SQUID microscopy 2252:Scanning SQUID microscope 2147: 2133: 2080: 2074:Scanning probe microscopy 1893:10.1007/s10854-011-0449-z 660:Scanning SQUID microscope 518:Scanning SQUID Microscope 49:scanning SQUID microscopy 2237:Scanning joule expansion 2232:Scanning ion-conductance 2217:Scanning electrochemical 2180:Magnetic resonance force 1332:Magnetic Current Imaging 439:between the electrodes, 213:(see figure). Above the 45:condensed matter physics 2288:Dip-pen nanolithography 1669:Applied Physics Letters 1361:Low-Temperature Physics 469:through the ring, and Φ 51:is a technique where a 2141: 1320: 1311: 1303: 1263: 1255: 1216: 1208: 1200: 1168: 1160: 1025: 919: 898: 865: 717: 698: 693:is the bias current, I 670:Earth's magnetic field 661: 528:) across an area. The 422: 202: 195: 175: 148: 121: 40: 2355:Measuring instruments 2242:Scanning Kelvin probe 2139: 1317: 1309: 1300: 1293:Short in a 3D Package 1280:Low-power radiography 1261: 1253: 1214: 1206: 1198: 1166: 1158: 1134:Printed Circuit Board 1026: 920: 893: 866: 704: 688: 659: 617:operating temperature 549:dilution refrigerator 545:helium-3 refrigerator 475:magnetic flux quantum 423: 196: 194:{\displaystyle \Phi } 176: 174:{\displaystyle I_{b}} 149: 147:{\displaystyle I_{a}} 122: 106: 35:sample stage. Right: 22: 2329:Vibrational analysis 2212:Scanning capacitance 1866:10.1143/JPSJ.61.4283 937: 909: 765: 583:thin-film deposition 228: 185: 158: 131: 111: 99:Operating principles 2227:Scanning Hall probe 2207:Piezoresponse force 2165:Electrostatic force 1858:1992JPSJ...61.4283S 1815:1994PhRvL..73..593T 1750:2015NatSR...5E8677W 1681:2000ApPhL..76.2304C 1638:1997ITAS....7.3134W 1578:1999ITAS....9.3704F 1478:1993ApPhL..62.2128B 1413:2015NatSR...515097S 1122:Through-Silicon Via 1118:Integrated Circuits 498:alternating current 284: 211:Josephson junctions 37:electron micrograph 31:. Bottom part is a 2170:Kelvin probe force 2142: 2115:Scanning tunneling 2026:2008-07-06 at the 1728:Scientific Reports 1501:The SQUID Handbook 1401:Scientific Reports 1321: 1312: 1304: 1264: 1256: 1217: 1209: 1201: 1169: 1161: 1037:spherical harmonic 1021: 915: 899: 861: 718: 699: 662: 631:e-beam lithography 568:magnetic shielding 534:integrated circuit 489:Φ < Φ 418: 416: 270: 203: 191: 171: 144: 117: 41: 2370:Superconductivity 2337: 2336: 1887:(10): 1602–1615. 1846:J. Phys. Soc. Jpn 1758:10.1038/srep08677 1646:10.1109/77.621996 1586:10.1109/77.783833 1472:(17): 2128–2130. 1421:10.1038/srep15097 1130:Multi-Chip Module 1126:System in package 918:{\displaystyle k} 855: 841: 826: 806: 780: 506:lock-in amplifier 408: 377: 308: 285: 253: 120:{\displaystyle I} 2377: 2350:Josephson effect 2298:Millipede memory 2267:Scanning voltage 2262:Scanning thermal 2067: 2060: 2053: 2044: 2043: 2002: 1999: 1993: 1990: 1984: 1980: 1974: 1971: 1965: 1962: 1956: 1955: 1953: 1947:. Archived from 1930: 1919: 1913: 1912: 1876: 1870: 1869: 1841: 1835: 1834: 1786: 1780: 1779: 1769: 1743: 1707: 1701: 1700: 1689:10.1063/1.126327 1664: 1658: 1657: 1621: 1615: 1612: 1606: 1605: 1561: 1555: 1552: 1546: 1542: 1536: 1533: 1527: 1524: 1515: 1514: 1496: 1490: 1489: 1486:10.1063/1.109448 1466:Appl. Phys. Lett 1449: 1443: 1442: 1432: 1392: 1371:Failure analysis 1351:Josephson Effect 1247: 1246: 1242: 1192: 1191: 1187: 1152: 1151: 1147: 1030: 1028: 1027: 1022: 1011: 1010: 989: 988: 964: 963: 924: 922: 921: 916: 895:Quantum vortices 870: 868: 867: 862: 856: 854: 853: 844: 843: 842: 834: 828: 827: 819: 809: 807: 805: 797: 796: 787: 782: 781: 773: 553:thin-film growth 427: 425: 424: 419: 417: 410: 409: 401: 399: 395: 394: 393: 388: 384: 383: 379: 378: 376: 375: 363: 347: 346: 325: 324: 309: 301: 293: 286: 283: 278: 266: 265: 256: 254: 246: 215:critical current 200: 198: 197: 192: 180: 178: 177: 172: 170: 169: 153: 151: 150: 145: 143: 142: 126: 124: 123: 118: 2385: 2384: 2380: 2379: 2378: 2376: 2375: 2374: 2340: 2339: 2338: 2333: 2302: 2271: 2197:Photon scanning 2143: 2131: 2120:Electrochemical 2108:Photoconductive 2076: 2071: 2028:Wayback Machine 2011: 2006: 2005: 2000: 1996: 1991: 1987: 1981: 1977: 1972: 1968: 1963: 1959: 1951: 1928: 1920: 1916: 1877: 1873: 1842: 1838: 1803:Phys. Rev. Lett 1800: 1796: 1792: 1787: 1783: 1723: 1719: 1715: 1708: 1704: 1665: 1661: 1622: 1618: 1613: 1609: 1562: 1558: 1553: 1549: 1543: 1539: 1534: 1530: 1525: 1518: 1511: 1497: 1493: 1463: 1459: 1455: 1450: 1446: 1393: 1389: 1384: 1347: 1334: 1295: 1282: 1273: 1248: 1244: 1240: 1238: 1237: 1193: 1189: 1185: 1183: 1182: 1153: 1149: 1145: 1143: 1142: 1100: 1089: 1085: 1035: = 2 1006: 1002: 984: 980: 959: 955: 938: 935: 934: 927:order parameter 910: 907: 906: 888: 876: 849: 845: 833: 832: 818: 817: 810: 808: 798: 792: 788: 786: 772: 771: 766: 763: 762: 755: 748: 744: 740: 736: 732: 716: 712: 708: 696: 692: 683: 654: 613: 606: 601:, for its high 595:thermal cycling 576:Meissner effect 564: 557:liquid nitrogen 522:magnetic fields 514: 512:Instrumentation 503: 492: 472: 463: 454:is the maximum 453: 415: 414: 400: 389: 371: 367: 362: 358: 354: 342: 338: 334: 330: 329: 320: 316: 315: 311: 310: 300: 291: 290: 279: 274: 261: 257: 255: 245: 238: 231: 229: 226: 225: 186: 183: 182: 165: 161: 159: 156: 155: 138: 134: 132: 129: 128: 112: 109: 108: 101: 17: 12: 11: 5: 2383: 2373: 2372: 2367: 2362: 2357: 2352: 2335: 2334: 2332: 2331: 2326: 2321: 2316: 2314:Nanotechnology 2310: 2308: 2304: 2303: 2301: 2300: 2295: 2290: 2285: 2279: 2277: 2273: 2272: 2270: 2269: 2264: 2259: 2254: 2249: 2244: 2239: 2234: 2229: 2224: 2219: 2214: 2209: 2204: 2199: 2194: 2193: 2192: 2182: 2177: 2175:Magnetic force 2172: 2167: 2162: 2160:Chemical force 2157: 2151: 2149: 2145: 2144: 2134: 2132: 2130: 2129: 2128: 2127: 2125:Spin polarized 2122: 2112: 2111: 2110: 2105: 2100: 2095: 2084: 2082: 2078: 2077: 2070: 2069: 2062: 2055: 2047: 2041: 2040: 2035: 2030: 2018: 2010: 2009:External links 2007: 2004: 2003: 1994: 1985: 1975: 1966: 1957: 1954:on 2019-02-20. 1914: 1871: 1836: 1809:(4): 593–596. 1798: 1794: 1790: 1781: 1721: 1717: 1713: 1702: 1659: 1616: 1607: 1556: 1547: 1537: 1528: 1516: 1509: 1491: 1461: 1457: 1453: 1444: 1386: 1385: 1383: 1380: 1379: 1378: 1373: 1368: 1363: 1358: 1353: 1346: 1343: 1333: 1330: 1294: 1291: 1281: 1278: 1272: 1269: 1236: 1233: 1181: 1178: 1141: 1138: 1098: 1087: 1083: 1020: 1017: 1014: 1009: 1005: 1001: 998: 995: 992: 987: 983: 979: 976: 973: 970: 967: 962: 958: 954: 951: 948: 945: 942: 914: 887: 884: 879: 878: 874: 871: 860: 852: 848: 840: 837: 831: 825: 822: 816: 813: 804: 801: 795: 791: 785: 779: 776: 770: 754: 751: 746: 742: 738: 734: 730: 714: 710: 706: 694: 690: 682: 679: 653: 650: 642:stepping motor 611: 604: 562: 513: 510: 501: 490: 470: 461: 451: 429: 428: 413: 407: 404: 398: 392: 387: 382: 374: 370: 366: 361: 357: 353: 350: 345: 341: 337: 333: 328: 323: 319: 314: 307: 304: 299: 296: 294: 292: 289: 282: 277: 273: 269: 264: 260: 252: 249: 244: 241: 239: 237: 234: 233: 190: 168: 164: 141: 137: 116: 100: 97: 59:strength with 57:magnetic field 15: 9: 6: 4: 3: 2: 2382: 2371: 2368: 2366: 2363: 2361: 2358: 2356: 2353: 2351: 2348: 2347: 2345: 2330: 2327: 2325: 2322: 2320: 2317: 2315: 2312: 2311: 2309: 2305: 2299: 2296: 2294: 2291: 2289: 2286: 2284: 2281: 2280: 2278: 2274: 2268: 2265: 2263: 2260: 2258: 2255: 2253: 2250: 2248: 2245: 2243: 2240: 2238: 2235: 2233: 2230: 2228: 2225: 2223: 2222:Scanning gate 2220: 2218: 2215: 2213: 2210: 2208: 2205: 2203: 2200: 2198: 2195: 2191: 2188: 2187: 2186: 2183: 2181: 2178: 2176: 2173: 2171: 2168: 2166: 2163: 2161: 2158: 2156: 2153: 2152: 2150: 2146: 2138: 2126: 2123: 2121: 2118: 2117: 2116: 2113: 2109: 2106: 2104: 2101: 2099: 2096: 2094: 2091: 2090: 2089: 2086: 2085: 2083: 2079: 2075: 2068: 2063: 2061: 2056: 2054: 2049: 2048: 2045: 2039: 2036: 2034: 2031: 2029: 2025: 2022: 2019: 2016: 2013: 2012: 1998: 1989: 1979: 1970: 1961: 1950: 1946: 1942: 1938: 1934: 1931:. IRPS 2004. 1927: 1926: 1918: 1910: 1906: 1902: 1898: 1894: 1890: 1886: 1882: 1875: 1867: 1863: 1859: 1855: 1851: 1847: 1840: 1832: 1828: 1824: 1820: 1816: 1812: 1808: 1804: 1785: 1777: 1773: 1768: 1763: 1759: 1755: 1751: 1747: 1742: 1737: 1733: 1729: 1725: 1706: 1698: 1694: 1690: 1686: 1682: 1678: 1674: 1670: 1663: 1655: 1651: 1647: 1643: 1639: 1635: 1631: 1627: 1620: 1611: 1603: 1599: 1595: 1591: 1587: 1583: 1579: 1575: 1571: 1567: 1560: 1551: 1541: 1532: 1523: 1521: 1512: 1510:3-527-40229-2 1506: 1502: 1495: 1487: 1483: 1479: 1475: 1471: 1467: 1448: 1440: 1436: 1431: 1426: 1422: 1418: 1414: 1410: 1406: 1402: 1398: 1391: 1387: 1377: 1376:Semiconductor 1374: 1372: 1369: 1367: 1364: 1362: 1359: 1357: 1354: 1352: 1349: 1348: 1342: 1338: 1329: 1325: 1316: 1308: 1299: 1290: 1286: 1277: 1268: 1260: 1252: 1243: 1232: 1229: 1225: 1221: 1213: 1205: 1197: 1188: 1177: 1173: 1165: 1157: 1148: 1137: 1135: 1131: 1127: 1123: 1119: 1115: 1110: 1108: 1104: 1096: 1091: 1081: 1077: 1073: 1069: 1065: 1060: 1057: 1053: 1048: 1046: 1042: 1038: 1034: 1012: 1007: 1003: 996: 990: 985: 981: 974: 971: 968: 960: 952: 946: 932: 928: 912: 904: 896: 892: 883: 872: 858: 850: 846: 835: 829: 820: 814: 811: 802: 799: 793: 789: 783: 774: 768: 761: 760: 759: 750: 726: 723: 703: 687: 678: 675: 671: 667: 658: 649: 647: 643: 639: 634: 632: 627: 621: 618: 614: 607: 600: 596: 592: 588: 584: 579: 577: 573: 569: 565: 558: 554: 550: 546: 542: 541:liquid helium 537: 535: 531: 527: 523: 519: 509: 507: 499: 494: 487: 483: 478: 476: 468: 467:magnetic flux 464: 457: 450: 446: 442: 438: 434: 411: 405: 402: 396: 390: 385: 380: 372: 359: 355: 351: 348: 343: 339: 335: 331: 326: 321: 317: 312: 305: 302: 297: 295: 287: 280: 275: 271: 267: 262: 258: 250: 247: 242: 240: 235: 224: 223: 222: 220: 216: 212: 208: 166: 162: 139: 135: 114: 105: 96: 94: 90: 86: 82: 78: 74: 70: 66: 62: 58: 54: 50: 46: 38: 34: 33:piezoelectric 30: 26: 21: 2276:Applications 2256: 2088:Atomic force 2015:John Kirtley 1997: 1988: 1978: 1969: 1960: 1949:the original 1924: 1917: 1884: 1880: 1874: 1852:(12): 4283. 1849: 1845: 1839: 1806: 1802: 1784: 1731: 1727: 1705: 1672: 1668: 1662: 1629: 1625: 1619: 1610: 1569: 1565: 1559: 1550: 1540: 1531: 1500: 1494: 1469: 1465: 1447: 1404: 1400: 1390: 1339: 1335: 1326: 1322: 1287: 1283: 1274: 1265: 1230: 1226: 1222: 1218: 1174: 1170: 1111: 1106: 1102: 1094: 1092: 1079: 1075: 1071: 1067: 1063: 1061: 1056:Bragg planes 1051: 1049: 1040: 1032: 900: 886:Applications 880: 756: 727: 719: 709:and (n+1/2)Φ 663: 635: 622: 609: 602: 580: 560: 538: 517: 515: 495: 486:sinusoidally 479: 459: 456:supercurrent 448: 440: 432: 430: 204: 76: 48: 42: 2103:Non-contact 2038:Neocera LLC 1114:mouse bites 587:lithography 95:compounds. 83:in several 73:lithography 29:tuning fork 2360:Microscopy 2344:Categories 2324:Microscopy 2319:Microscope 2093:Conductive 1741:1807.06746 1382:References 1356:BCS theory 1120:(IC) with 931:energy gap 638:cantilever 626:inductance 530:microscope 482:transducer 437:resistance 87:including 65:resolution 61:micrometre 2190:Nano-FTIR 1909:136660553 1901:0957-4522 1799:7−δ 1722:7−x 1697:0003-6951 1654:1051-8223 1594:1051-8223 1407:: 15097. 997:− 957:Δ 941:Δ 903:isotropic 839:→ 830:× 824:→ 803:π 790:μ 778:→ 722:rastering 681:Operation 646:vibration 369:Φ 365:Φ 360:π 352:⁡ 327:− 268:− 189:Φ 2307:See also 2098:Infrared 2024:Archived 1831:10057486 1776:25728772 1734:: 8677. 1602:25032903 1439:26459874 1345:See also 1105:, where 572:mu-metal 69:rastered 25:helium-4 1945:2307126 1854:Bibcode 1811:Bibcode 1767:4345321 1746:Bibcode 1677:Bibcode 1634:Bibcode 1574:Bibcode 1474:Bibcode 1430:4602221 1409:Bibcode 1136:(PCB). 1128:(SiP), 1124:(TSV), 591:Niobium 473:is the 445:current 443:is the 435:is the 219:voltage 63:-scale 2081:Common 1983:69-76. 1943:  1907:  1899:  1829:  1774:  1764:  1695:  1652:  1600:  1592:  1507:  1437:  1427:  1319:plane. 1302:paper. 1239:": --> 1184:": --> 1144:": --> 1095:et al. 1093:Tsuei 1052:et al. 597:, and 431:where 77:et al. 2148:Other 1952:(PDF) 1941:S2CID 1929:(PDF) 1905:S2CID 1736:arXiv 1598:S2CID 1366:SQUID 1045:phase 526:SQUID 93:BSCCO 1897:ISSN 1827:PMID 1772:PMID 1693:ISSN 1650:ISSN 1590:ISSN 1505:ISBN 1435:PMID 1241:edit 1186:edit 1146:edit 1090:/2. 599:YBCO 154:and 91:and 89:YBCO 1933:doi 1889:doi 1862:doi 1819:doi 1801:". 1762:PMC 1754:doi 1685:doi 1642:doi 1582:doi 1482:doi 1425:PMC 1417:doi 578:). 547:or 349:cos 43:In 2346:: 1939:. 1903:. 1895:. 1885:22 1883:. 1860:. 1850:61 1848:. 1825:. 1817:. 1807:73 1805:. 1793:Cu 1770:. 1760:. 1752:. 1744:. 1730:. 1726:. 1716:Cu 1691:. 1683:. 1673:76 1671:. 1648:. 1640:. 1628:. 1596:. 1588:. 1580:. 1568:. 1519:^ 1480:. 1470:62 1468:. 1456:Cu 1433:. 1423:. 1415:. 1403:. 1399:. 1101:/2 733:Cu 674:nA 666:pT 516:A 477:. 458:, 447:, 207:DC 47:, 2066:e 2059:t 2052:v 1935:: 1911:. 1891:: 1868:. 1864:: 1856:: 1833:. 1821:: 1813:: 1797:O 1795:3 1791:2 1778:. 1756:: 1748:: 1738:: 1732:5 1720:O 1718:3 1714:2 1699:. 1687:: 1679:: 1656:. 1644:: 1636:: 1630:7 1604:. 1584:: 1576:: 1570:9 1513:. 1488:. 1484:: 1476:: 1462:7 1460:O 1458:3 1454:2 1441:. 1419:: 1411:: 1405:5 1245:] 1190:] 1150:] 1107:A 1103:A 1099:0 1088:0 1084:0 1080:n 1076:n 1072:n 1068:n 1064:n 1041:i 1033:l 1019:) 1016:) 1013:a 1008:y 1004:k 1000:( 994:) 991:a 986:x 982:k 978:( 975:s 972:o 969:c 966:( 961:0 953:= 950:) 947:k 944:( 913:k 875:0 859:. 851:3 847:r 836:r 821:l 815:d 812:I 800:4 794:0 784:= 775:B 769:d 747:0 743:o 739:7 737:O 735:3 731:2 715:0 711:0 707:0 695:0 691:b 612:c 610:T 605:c 603:T 563:c 561:T 502:0 491:0 471:0 462:c 460:I 452:0 449:I 441:I 433:R 412:, 406:2 403:1 397:) 391:2 386:) 381:) 373:0 356:( 344:c 340:I 336:2 332:( 322:2 318:I 313:( 306:2 303:R 298:= 288:, 281:2 276:0 272:I 263:2 259:I 251:2 248:R 243:= 236:V 167:b 163:I 140:a 136:I 115:I

Index


helium-4
tuning fork
piezoelectric
electron micrograph
condensed matter physics
superconducting quantum interference device
magnetic field
micrometre
resolution
rastered
lithography
unconventional superconductivity
high-temperature superconductors
YBCO
BSCCO

DC
Josephson junctions
critical current
voltage
resistance
current
supercurrent
magnetic flux
magnetic flux quantum
transducer
sinusoidally
alternating current
lock-in amplifier

Text is available under the Creative Commons Attribution-ShareAlike License. Additional terms may apply.